Vista de página 4
Model 2200D/P91 Page 3
TEMPERATURE RANGE:
0-40°C (32 to 104°F)
WETTED MATERIALS:
PVC and Teflon F.E.P.
FAILURE ALARM:
5A SPDF relay closes if membrane is punctured. LED on panel
is illuminated
CABLE:
Integral 8 m (26 ft.) Maximum distance to analyzer 300 m (1000
ft.)
REPLACEMENT CARTRIDGE:
Part No. P91-D. Hermetically sealed.
MOUNTING HARDWARE
FLOTATION MOUNTING:
Model MHBF consists of ball float, extension pipe, swivel
bracket, junction box and 50 feet of interconnect cable
SUBMERSION MOUNTING:
Model MHDS consists of 5 ft. PVC pipe, junction box and 50 feet
of interconnect cable
FLOW-THROUGH ASSEMBLY:
Model MHFT. Pipe connections are 2” NPT